Pacemakers in Cardiac Health

Pacemaker
By:
Somesh Kumar Malhotra
Assistant Professor,
ECE Deptt.,UIET,CSJM University
3/10/2025
1
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Introduction
The rhythmic action of the heart is intiated by
regularly recurring action potential
(electrochemical impulse) originating at the
natural cardiac pacemaker located at the
sinoartrial (SA) node.
Each pacing impulse is propogated throughout
the myocardium, spreading over the surface
of the artria to the artrio ventricular valve and
depolarizing the artria.
3/10/2025
2
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Introduction
After the brief delay at the AV node the
impulse is rapidly conducted to the ventricle
to depolarize the ventricular musculature.
 A normal sinus rhythm (NSR) depends on the
continuous, periodic performance of the
pacemaker and the integrity of the neuronal
conducting pathways.
Any change in the NSR is called an arrhythmia
(abnormal rhythm)
3/10/2025
3
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Introduction
Should the SA node temporarily or permanently
fail because of diseases (SA node diseases).
Under these condition the heart is paced at a
much slower rate than normal, ranging between
30 and 50 beats per minute(BPM).
The result is a condition called bradycardia (slow
heart), in which the heart cannot provide
sufficient blood circulation to meet the body’s
physical demands.
3/10/2025
4
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Introduction
During the transition period from a NSR to a slow
rhythm, dizziness and loss of consciousness
(syncope) may occur because of diminished
cardiac output.
 Heart block occurs whenever the conduction
system fails to transmits the pacing impulse from
the artria to ventricle properly.
First degree block 
an excessive delay at the AV
node occurs that causes the P-R interval to
exceed 0.2 seconds for normal adults.
3/10/2025
5
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Introduction
Second degree block 
results in the complete
but intermittent inhibition of the pacing
impulse, which may occur at the AV node.
Total and continuous impulse blockage is
called 
third degree block
.
In all these conditions ,an artificial method of
pacing is generally required to ensure that
the heart beat at the rate that is sufficient to
maintain the proper circulation.
3/10/2025
6
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker System
3/10/2025
7
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker System
3/10/2025
8
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker Components
Pulse Generator
Electronic Circuitry
Lead system
Pulse Generator
Lithium cell is the current standard battery
Advantages:
Long life – 4 to 10 years
Output voltage decreases gradually with time making
sudden battery failure unlikely
Pacemaker System
3/10/2025
9
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
 
Pacemaker System
3/10/2025
10
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
 
Pacemaker System
3/10/2025
11
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
 
Pacemaker System
3/10/2025
12
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
 
Pacemaker System
3/10/2025
13
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker System
3/10/2025
14
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker System
3/10/2025
15
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker System
3/10/2025
16
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker has two main functions:
Sense intrinsic cardiac electrical activity
Electrically stimulate the heart
VVI- senses intrinsic cardiac activity in the
ventricle and when a preset interval of time
with no ventricular activity occurs it
depolarizes the right ventricle causing
ventricular contraction
Pacemaker System
3/10/2025
17
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Dual chamber pacer is more complicated
because the pacer has the ability to both
sense and pace either the atrium or the
ventricle
Possible to have only atrial, only ventricular or
both atrial and ventricular pacing
DDD pacer is a common example of this
Pacemaker System
3/10/2025
18
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacemaker System
3/10/2025
19
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
 
Pacing Modes and Pulse Generator
3/10/2025
20
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Several pacing techniques with both internal
and external pacemaker.
They can be classified as either competitive
and noncompetitive pacing mode.
The first pulse generators were fixed rate or
asynchronous device that produced pulses at
a fixed rate and were independent of any
natural cardiac activity.
Pacing Modes and Pulse Generator
3/10/2025
21
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Pacing Modes and Pulse Generator
3/10/2025
22
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Asynchronous pacing is called 
competitive
pacing
 because the fixed rate impulse may
occur along with natural pacing impulse
generated by the heart and would therefore
be in competition with them in controlling the
heartbeat..
This competition is largely eliminated through
use of ventricular or artrial programmed pulse
generator.
Pacing Modes and Pulse Generator
3/10/2025
23
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
The ventricular programmed pulse generator,
when connected to the via. Electrodes, is able
to sense the presence (or absence) of
naturally occurring R wave .
The output of an R wave inhibited (demand)
unit is suppressed (no output pulse are
produced) as long as natural (intrinsic) R
waves are present
Pacing Modes and Pulse Generator
3/10/2025
24
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Sometime the pacemaker are preset for intrinsic rate
fall (around 70 BPM), the unit will automatically
provide an output to pace the heart after an escape
interval at the designated rate.
In this way ventricular inhibited pacers are able to pace
on demand, normally having different controls of
operation with constant rate  or somewhere between
30 and 180 BPM rate with amplitude of ouput pacing
pulse between 0.1 and 20 mA.
They also have indicators for each detected R wave or
pacer initiated pulse.
Pacing Modes and Pulse Generator
3/10/2025
25
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
When required, the basic pacing rate of some of
the earlier implanted pacers may be changed
with the use of needle like screw driver (a Keith
surgical skin needle) that is inserted to alter the
rate control in the pulse generator.
The amplitude of the impulse may also adjusted
in some earlier pacers by using the same type of
needle in the appropriate control.
Pacing Modes and Pulse Generator
3/10/2025
26
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
In a newer type of pacer, these adjustments are
accomplished by means of coded impulses that are
magnetically coupled to the implanted pulse generator
from the skin surface, thus eliminating the need to
puncture the skin.
To adjust the pacemaker, a special programming
device with an attached coil is placed over the planted
pulse generator.
Appropriate controls of the programmer allows the
unit to transmit coded signals that cause the pacer to
change its basic rate and vary the amplitude of the
pulses.
Pacing Modes and Pulse Generator
3/10/2025
27
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Similarly R wave triggered pulse generator
works, only basic difference is the unit is
triggered rather than inhibited by each R
wave.
Ventricular triggered pacing is used less
frequently than inhibited mode pacing.
Similarly the 
artrial programmed  pacer 
are
responsible for synchronisation of P-wave
Pacing Modes and Pulse Generator
3/10/2025
28
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Fig below shows a block diagram showing the
components of the circuitry.
The timing circuit which consist of an RC
network, a reference voltage source , and a
comparator determines the basic pacing rate
of the pulse generator.
Its output signal is fed to second RC network,
the pulse width circuit, which determine the
pulse duration.
Pacing Modes and Pulse Generator
3/10/2025
29
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
A third RC circuit network, the rate limiting circuit
,disables the the comparator for a preset  interval
and thus limits the pacing rate to a maximum of
120 pulses per minutes.
The output circuit provides a voltage pulse to
simulate the heart.
The voltage monitor circuit sense cell depletion
and signals the rate slow down circuit and energy
compensation circuit of this event
Pacing Modes and Pulse Generator
3/10/2025
30
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
The rate slow down the circuit shuts off some
of the current to the basic timing network to
cause the the rate to slow down 8 +/- 3 beats
per minute when cell depletion has occcured.
The energy compensation circuit causes the
pulse duration to increase as the battery
voltage decreases, to maintain nearly constant
simulation energy to the heart.
Pacing Modes and Pulse Generator
3/10/2025
31
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
There is also the feedback loop from output circuit to
the refractory circuit  which provides a period of time
following an output pulse or a sense R-wave during
which the amplifier will not respond to outside signals.
The sensing circuit detect a spontaneous R wave and
reset the oscillator timing capacitor.
The reversion circuit allows the amplifier to detect a
spontaneous R wave in the presence of low level
continuous wave interference. In the absence of an R
wave, the circuit allows the oscillator to pace at its rate
+/- 1 beat per minute.
Pacing Modes and Pulse Generator
3/10/2025
32
by Er.Somesh Kr Malhotra,ECE
Deptt,UIET,CSJM University,Kanpur
Slide Note
Embed
Share

Initiation of heart rhythm, implications of pacemaker failure, and the necessity of artificial pacing methods for maintaining proper circulation are discussed in this informative content presented by Er. Somesh Kr. Malhotra, ECE Dept., UIET, CSJM University, Kanpur.

  • Pacemakers
  • Cardiac Health
  • Heart Rhythm
  • Artificial Pacing
  • Circulation

Uploaded on Mar 10, 2025 | 0 Views


Download Presentation

Please find below an Image/Link to download the presentation.

The content on the website is provided AS IS for your information and personal use only. It may not be sold, licensed, or shared on other websites without obtaining consent from the author.If you encounter any issues during the download, it is possible that the publisher has removed the file from their server.

You are allowed to download the files provided on this website for personal or commercial use, subject to the condition that they are used lawfully. All files are the property of their respective owners.

The content on the website is provided AS IS for your information and personal use only. It may not be sold, licensed, or shared on other websites without obtaining consent from the author.

E N D

Presentation Transcript


  1. Pacemaker By: Somesh Kumar Malhotra Assistant Professor, ECE Deptt.,UIET,CSJM University by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 1

  2. Introduction The rhythmic action of the heart is intiated by regularly recurring (electrochemical impulse) originating at the natural cardiac pacemaker located at the sinoartrial (SA) node. Each pacing impulse is propogated throughout the myocardium, spreading over the surface of the artria to the artrio ventricular valve and depolarizing the artria. action potential by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 2

  3. Introduction After the brief delay at the AV node the impulse is rapidly conducted to the ventricle to depolarize the ventricular musculature. A normal sinus rhythm (NSR) depends on the continuous, periodic performance of the pacemaker and the integrity of the neuronal conducting pathways. Any change in the NSR is called an arrhythmia (abnormal rhythm) by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 3

  4. Introduction Should the SA node temporarily or permanently fail because of diseases (SA node diseases). Under these condition the heart is paced at a much slower rate than normal, ranging between 30 and 50 beats per minute(BPM). The result is a condition called bradycardia (slow heart), in which the heart cannot provide sufficient blood circulation to meet the body s physical demands. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 4

  5. Introduction During the transition period from a NSR to a slow rhythm, dizziness and loss of consciousness (syncope) may occur because of diminished cardiac output. Heart block occurs whenever the conduction system fails to transmits the pacing impulse from the artria to ventricle properly. First degree block an excessive delay at the AV node occurs that causes the P-R interval to exceed 0.2 seconds for normal adults. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 5

  6. Introduction Second degree block results in the complete but intermittent inhibition of the pacing impulse, which may occur at the AV node. Total and continuous impulse blockage is called third degree block. In all these conditions ,an artificial method of pacing is generally required to ensure that the heart beat at the rate that is sufficient to maintain the proper circulation. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 6

  7. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 7

  8. Pacemaker System Pacemaker Components Pulse Generator Electronic Circuitry Lead system Pulse Generator Lithium cell is the current standard battery Advantages: Long life 4 to 10 years Output voltage decreases gradually with time making sudden battery failure unlikely by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 8

  9. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 9

  10. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 10

  11. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 11

  12. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 12

  13. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 13

  14. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 14

  15. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 15

  16. Pacemaker System Pacemaker has two main functions: Sense intrinsic cardiac electrical activity Electrically stimulate the heart VVI- senses intrinsic cardiac activity in the ventricle and when a preset interval of time with no ventricular activity occurs it depolarizes the right ventricle causing ventricular contraction by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 16

  17. Pacemaker System Dual chamber pacer is more complicated because the pacer has the ability to both sense and pace either the atrium or the ventricle Possible to have only atrial, only ventricular or both atrial and ventricular pacing DDD pacer is a common example of this by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 17

  18. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 18

  19. Pacemaker System by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 19

  20. Pacing Modes and Pulse Generator Several pacing techniques with both internal and external pacemaker. They can be classified as either competitive and noncompetitive pacing mode. The first pulse generators were fixed rate or asynchronous device that produced pulses at a fixed rate and were independent of any natural cardiac activity. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 20

  21. Pacing Modes and Pulse Generator by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 21

  22. Pacing Modes and Pulse Generator Asynchronous pacing is called competitive pacing because the fixed rate impulse may occur along with natural pacing impulse generated by the heart and would therefore be in competition with them in controlling the heartbeat.. This competition is largely eliminated through use of ventricular or artrial programmed pulse generator. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 22

  23. Pacing Modes and Pulse Generator The ventricular programmed pulse generator, when connected to the via. Electrodes, is able to sense the presence (or absence) of naturally occurring R wave . The output of an R wave inhibited (demand) unit is suppressed (no output pulse are produced) as long as natural (intrinsic) R waves are present by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 23

  24. Pacing Modes and Pulse Generator Sometime the pacemaker are preset for intrinsic rate fall (around 70 BPM), the unit will automatically provide an output to pace the heart after an escape interval at the designated rate. In this way ventricular inhibited pacers are able to pace on demand, normally having different controls of operation with constant rate or somewhere between 30 and 180 BPM rate with amplitude of ouput pacing pulse between 0.1 and 20 mA. They also have indicators for each detected R wave or pacer initiated pulse. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 24

  25. Pacing Modes and Pulse Generator When required, the basic pacing rate of some of the earlier implanted pacers may be changed with the use of needle like screw driver (a Keith surgical skin needle) that is inserted to alter the rate control in the pulse generator. The amplitude of the impulse may also adjusted in some earlier pacers by using the same type of needle in the appropriate control. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 25

  26. Pacing Modes and Pulse Generator In a newer type of pacer, these adjustments are accomplished by means of coded impulses that are magnetically coupled to the implanted pulse generator from the skin surface, thus eliminating the need to puncture the skin. To adjust the pacemaker, a special programming device with an attached coil is placed over the planted pulse generator. Appropriate controls of the programmer allows the unit to transmit coded signals that cause the pacer to change its basic rate and vary the amplitude of the pulses. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 26

  27. Pacing Modes and Pulse Generator Similarly R wave triggered pulse generator works, only basic difference is the unit is triggered rather than inhibited by each R wave. Ventricular triggered pacing is used less frequently than inhibited mode pacing. Similarly the artrial programmed pacer are responsible for synchronisation of P-wave by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 27

  28. Pacing Modes and Pulse Generator Fig below shows a block diagram showing the components of the circuitry. The timing circuit which consist of an RC network, a reference voltage source , and a comparator determines the basic pacing rate of the pulse generator. Its output signal is fed to second RC network, the pulse width circuit, which determine the pulse duration. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 28

  29. Pacing Modes and Pulse Generator A third RC circuit network, the rate limiting circuit ,disables the the comparator for a preset interval and thus limits the pacing rate to a maximum of 120 pulses per minutes. The output circuit provides a voltage pulse to simulate the heart. The voltage monitor circuit sense cell depletion and signals the rate slow down circuit and energy compensation circuit of this event by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 29

  30. Pacing Modes and Pulse Generator The rate slow down the circuit shuts off some of the current to the basic timing network to cause the the rate to slow down 8 +/- 3 beats per minute when cell depletion has occcured. The energy compensation circuit causes the pulse duration to increase as the battery voltage decreases, to maintain nearly constant simulation energy to the heart. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 30

  31. Pacing Modes and Pulse Generator There is also the feedback loop from output circuit to the refractory circuit which provides a period of time following an output pulse or a sense R-wave during which the amplifier will not respond to outside signals. The sensing circuit detect a spontaneous R wave and reset the oscillator timing capacitor. The reversion circuit allows the amplifier to detect a spontaneous R wave in the presence of low level continuous wave interference. In the absence of an R wave, the circuit allows the oscillator to pace at its rate +/- 1 beat per minute. by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 31

  32. Pacing Modes and Pulse Generator by Er.Somesh Kr Malhotra,ECE Deptt,UIET,CSJM University,Kanpur 3/10/2025 32

Related


More Related Content

giItT1WQy@!-/#giItT1WQy@!-/#giItT1WQy@!-/#