Latest Update on Sci-ECAL MIP Results by Jiaxuan Wang

Slide Note
Embed
Share

Jiaxuan Wang presents an update on Sci-ECAL MIP results as of March 8, 2024. The new method for muon energy reconstruction involves multiple fittings to exclude noise hits and obtain the real track and hits. The muon energy is estimated around 0.3049 MeV using new MIP calibration results. Despite progress, there are still abnormal channels and dead channels with fit problems, resulting in 9.26% total dead channels.


Uploaded on Sep 18, 2024 | 0 Views


Download Presentation

Please find below an Image/Link to download the presentation.

The content on the website is provided AS IS for your information and personal use only. It may not be sold, licensed, or shared on other websites without obtaining consent from the author. Download presentation by click this link. If you encounter any issues during the download, it is possible that the publisher has removed the file from their server.

E N D

Presentation Transcript


  1. Sci-ECAL MIP results update 2024/03/08 Jiaxuan Wang

  2. Motivation Old method: center of gravity method, and fit once Landau MPV from MIP spectrum Muon energy reconstruction in channel level New method: fit multi times Data selection : center of gravity method To exclude the noise hit far away from real muon track(more than 10 strips far away) First fitting: To exclude the noise hit near the real muon track(2 strips far away) Second fitting: To exclude the noise not on the track Third fitting: Get the real track and real hits 0.305MeV Blue : All channels Red : Dead Channels MeV 2

  3. Muon energy reconstruction Muon energy using new MIP calibrated results Mean value around 0.3049 MeV MeV 0.305MeV Blue : All channels Red : Dead Channels 3 MeV MeV

  4. Abnormal channels Still some abnormal channels Dead-channel list update MeV Inadequate statistics Fit problem Total dead channels Percentage 9.26% 5.03% 14.29% Channels 584 317 901 4

Related


More Related Content